Understanding Repeatability in Packaging Dimension Measurement
Repeated readings of the same packaging dimension can vary. If the random contributions are independent, have a common variance and are centered, the standard deviation of the mean from that component falls in proportion to one divided by the square root of the number of readings. The spread of the individual readings is unchanged by taking the average.
Drift and other shared conditions can make repeated readings correlated. Evaluate those effects separately rather than applying the same reduction to all uncertainty components.
Why Calibration Error Does Not Average Away
A shared offset shifts each reading and therefore shifts their mean by the same amount. Correct a recognized systematic effect when appropriate, then include the uncertainty of the correction in the reported result.
The calibration correction can be common to a series of readings taken with the same instrument. Its uncertainty does not reduce just because there are more readings in that series.
A Hypothetical Cosmetic Packaging Example
For a hypothetical lid diameter, suppose the independent repeatability standard deviation is 0.020 mm. The independent component of the mean from 16 readings is 0.005 mm. If the same unknown +0.010 mm instrument offset applies to every reading in the illustration, the mean retains that +0.010 mm offset. A suitable reference check can help characterize an offset, while additional repeats cannot separate it from the part dimension.